"Regarding the cartridge, we found a technical problem with an uneven damping unit. This existed most probably already partly before delivery, but since all values were within the tolerated area, it did not call attention. It seems as the problem increased by transport / time because when we received the cartridge back, it was not possible to achieve correct values with this damping unit. We have changed the damping unit for a new one and re-adjusted the cartridge."
全新唱頭借給我測試,發現數據遠不如預期。寄回原廠調整後,數據如下,有明顯改善。上面是原廠說明做了些什麼。
After the adjustments, the channel separation measures much better in the 35dB range. Below are the before and after frequency response measurements:
The square wave test results, right channel. The stylus resonance is now better controlled.